Patent No. 6,089,098
|
SPECIFICATIONS
Service: Air and compatible combustible gases.
Wetted Materials: Consult factory.
Temperature Limits: 0 to 140°F (-17 to 60°C). Note:
Set point drift may occur with ambient temperature changes.
Pressure Limits: 45" w.c. (11.2 kPa) continuous; 10 psig (68.95
kPa) surge.
Enclosure Rating: IP54, NEMA 3, 7 and 9.
Switch Type: 1 Form C relay (SPDT).
Electrical Rating: 10A, 120/240 VAC, 28 VDC. Resistive 50mA,
125 VDC.
Power Requirements: 24 VDC ±10%. 120 or 240 VAC ±10%
optional.
Electrical Connections: Internal terminal block.
Process Connections: 1/8" female NPT.
Mounting Orientation: Diaphragm in vertical position. Consult
factory for other position orientations.
Set Point Adjustment: Screw type on top of housing.
Weight: 2 lbs. 15.7 oz. (1.35 kg).
Agency Approvals: CE, UL, CSA, FM, ATEX. |